% IMPORTANT: The following is UTF-8 encoded.  This means that in the presence
% of non-ASCII characters, it will not work with BibTeX 0.99 or older.
% Instead, you should use an up-to-date BibTeX implementation like “bibtex8” or
% “biber”.

@ARTICLE{Mikirtychyants:138437,
      author       = {Mikirtychyants, M. and Engels, Ralf W. and Grigoryev, K.
                      and Kleines, H. and Kravtsov, P. and Lorenz, S. and
                      Nekipelov, M. and Nelyubin, V. and Rathmann, F. and Sarkadi,
                      J. and Paetz gen. Schieck, H. and Seyfarth, H. and Steffens,
                      E. and Stroeher, Hans and Vasilyev, A.},
      title        = {{T}he polarized {H} and {D} atomic beam source for {ANKE}
                      at {COSY}-{J}ülich},
      journal      = {Nuclear instruments $\&$ methods in physics research / A},
      volume       = {721},
      issn         = {0168-9002},
      address      = {Amsterdam},
      publisher    = {North-Holland Publ. Co.},
      reportid     = {FZJ-2013-04568},
      pages        = {83 - 98},
      year         = {2013},
      note         = {18 pages, 26 figures, 4 tables},
      abstract     = {A polarized atomic beam source was developed for the
                      polarized internal storage-cell gas target at the magnet
                      spectrometer ANKE of $COSY-J\'ulich.$ The intensities of the
                      beams injected into the storage cell, measured with a
                      compression tube, are $7.5\cdot 10^{16}$ hydrogen atoms/s
                      (two hyperfine states) and $3.9\cdot 10^{16}$ deuterium
                      atoms/s (three hyperfine states). For the hydrogen beam the
                      achieved vector polarizations are $p_{\rm z}\approx\pm0.92$.
                      For the deuterium beam, the obtained combinations of vector
                      and tensor ($p_{\rm zz}$) polarizations are $p_{\rm
                      z}\approx\pm 0.90$ (with a constant $p_{\rm zz}\approx
                      +0.86$), and $p_{\rm zz}=+0.90$ or $p_{\rm zz}=-1.71$ (both
                      with vanishing $p_{\rm z}$). The paper includes a detailed
                      technical description of the apparatus and of the
                      investigations performed during the development.},
      cin          = {IKP-2 / ZEA-2 / IKP-TA},
      ddc          = {530},
      cid          = {I:(DE-Juel1)VDB24 / I:(DE-Juel1)ZEA-2-20090406 /
                      I:(DE-Juel1)IKP-TA-20111104},
      pnm          = {531 - Hadron Structure and Dynamics (HSD) (POF2-531)},
      pid          = {G:(DE-HGF)POF2-531},
      typ          = {PUB:(DE-HGF)16},
      eprint       = {1212.1840},
      howpublished = {arXiv:1212.1840},
      archivePrefix = {arXiv},
      SLACcitation = {$\%\%CITATION$ = $arXiv:1212.1840;\%\%$},
      UT           = {WOS:000320598500013},
      doi          = {10.1016/j.nima.2013.03.043},
      url          = {https://juser.fz-juelich.de/record/138437},
}