@ARTICLE{Boothroyd:201611,
author = {Boothroyd, C. B. and Kasama, T. and Dunin-Borkowski, Rafal},
title = {{C}omparison of approaches and artefacts in the measurement
of detector modulation transfer functions},
journal = {Ultramicroscopy},
volume = {129},
issn = {0304-3991},
address = {Amsterdam},
publisher = {Elsevier Science},
reportid = {FZJ-2015-03905},
pages = {18 - 29},
year = {2013},
abstract = {In order to investigate the reproducibility of measurements
of transmission electron microscope detector modulation
transfer functions (MTFs) we measure the MTF of a
charge-coupled device (CCD) camera using five different
methods. MTFs derived from a sharp edge, a circular aperture
and electron holographic interference fringes are found to
agree closely with one other. The difficulty of obtaining
accurate measurements of MTFs and the potential of using
focused electron probes to make direct measurements of MTFs
is discussed. We highlight the sensitivity of image contrast
after deconvolution to small differences in the measured
MTF.},
cin = {PGI-5},
ddc = {570},
cid = {I:(DE-Juel1)PGI-5-20110106},
pnm = {424 - Exploratory materials and phenomena (POF2-424)},
pid = {G:(DE-HGF)POF2-424},
typ = {PUB:(DE-HGF)16},
UT = {WOS:000318615800003},
doi = {10.1016/j.ultramic.2013.03.001},
url = {https://juser.fz-juelich.de/record/201611},
}