abstract = {The irradiation facility at an old medical cyclotron (Ep=17
MeV; Ed=10 MeV) was upgraded by extending the beam line and
incorporation of solid state targetry. Tests performed to
check the quality of the available beam are outlined.
Results on nuclear data measurements and improvement of
radiochemical separations are described. Using solid
targets, with the proton beam falling at a slanting angle of
20°, a few radionuclides, e.g. 75Se, 120I, 124I, etc. were
produced with medium currents (up to 20 µA) in
no-carrier-added form in quantities sufficient for local
use. The extended irradiation facility has considerably
enhanced the utility of the medical cyclotron.Keywords:
Medical cyclotron; Extended beam line; Solid state targetry;
Nuclear data measurement; Activation of materials;
Production of radionuclides},