% IMPORTANT: The following is UTF-8 encoded.  This means that in the presence
% of non-ASCII characters, it will not work with BibTeX 0.99 or older.
% Instead, you should use an up-to-date BibTeX implementation like “bibtex8” or
% “biber”.

@ARTICLE{Ibach:22162,
      author       = {Ibach, H. and Rajeswari, J.},
      title        = {{E}lectron energy loss spectrometers: {A}n advanced
                      operation mode for the lens system and the quantitative
                      calculation of solid angle and transmission},
      journal      = {Journal of electron spectroscopy and related phenomena},
      volume       = {185},
      number       = {3-4},
      issn         = {0368-2048},
      address      = {New York, NY [u.a.]},
      publisher    = {Elsevier},
      reportid     = {PreJuSER-22162},
      pages        = {61–70},
      year         = {2012},
      note         = {One of the authors (J.R.) gratefully acknowledges the
                      financial support from NRW Research School "Forschung mit
                      Synchrotron-strahlung in den Nano- und Biowissenschaften".},
      abstract     = {A new operation mode of the lens system in electron energy
                      loss spectrometers is proposed which brings the cardinal
                      plane of the lenses closer to the sample. Experiments as
                      well as electron optical calculations show an intensity gain
                      by a factor of 3-4. The required voltages on the lens
                      elements are calculated as function of the impact energy at
                      the sample. Furthermore, the lens system is characterized
                      with respect to the solid angle probed by the spectrometer
                      and the transmission which enables the evaluation of
                      spectral intensities in terms of the absolute scattering
                      probability for elastic and inelastic scattering. The
                      results apply to most existing electron energy loss
                      spectrometers as they stem from the same design and use
                      identical lens systems. (C) 2012 Elsevier B.V. All rights
                      reserved.},
      keywords     = {J (WoSType)},
      cin          = {PGI-6 / PGI-3},
      ddc          = {620},
      cid          = {I:(DE-Juel1)PGI-6-20110106 / I:(DE-Juel1)PGI-3-20110106},
      pnm          = {Grundlagen für zukünftige Informationstechnologien},
      pid          = {G:(DE-Juel1)FUEK412},
      shelfmark    = {Spectroscopy},
      typ          = {PUB:(DE-HGF)16},
      UT           = {WOS:000304640700003},
      doi          = {10.1016/j.elspec.2012.01.001},
      url          = {https://juser.fz-juelich.de/record/22162},
}