TY  - JOUR
AU  - Schmalzl, K.
AU  - Schmidt, W.
AU  - Raymond, S.
AU  - Feilbach, H.
AU  - Mounier, C.
AU  - Vettard, B.
AU  - Brückel, T.
TI  - The upgrade of the cold neutron three-axis spectrometer IN12 at the ILL
JO  - Nuclear instruments & methods in physics research / A
VL  - 819
SN  - 0168-9002
CY  - Amsterdam
PB  - North-Holland Publ. Co.
M1  - FZJ-2016-01865
SP  - 89 - 98
PY  - 2016
AB  - After nearly 40 years of successful operation the cold three-axis spectrometer IN12 at the Institut Laue-Langevin, Grenoble, France, has been relocated to a new position and the primary spectrometer has been upgraded. Latest modern optical components are employed. A new guide in combination with a virtual source concept and a double focusing monochromator guarantee highest flux. With its high unpolarized and polarized neutron flux IN12 allows for demanding experiments. A velocity selector in the guide ensures a clean beam and a very low background. A gain in flux of about an order of magnitude at the sample position has been achieved compared to the previous instrument and IN12׳s wavelength range now extends far into the warmish region.
LB  - PUB:(DE-HGF)16
UR  - <Go to ISI:>//WOS:000372318800013
DO  - DOI:10.1016/j.nima.2016.02.067
UR  - https://juser.fz-juelich.de/record/283549