% IMPORTANT: The following is UTF-8 encoded.  This means that in the presence
% of non-ASCII characters, it will not work with BibTeX 0.99 or older.
% Instead, you should use an up-to-date BibTeX implementation like “bibtex8” or
% “biber”.

@INPROCEEDINGS{Ptter:811981,
      author       = {Pütter, Sabine and Syed Mohd, Amir and Mattauch, Stefan
                      and Koutsioumpas, Alexandros and Schneider, Harald and
                      Brückel, Thomas},
      title        = {{H}ow to accomplish (quasi) in-situ neutron reflectivity
                      measurements on ultrathin magnetic films},
      reportid     = {FZJ-2016-04282},
      year         = {2016},
      abstract     = {The investigation of ultrathin films which are sensitive to
                      ambient air at large scale instruments is a challenge as the
                      capabilities for on-site growth UHV chambers to measure
                      in-situ are often limited, mostly due to lacking space. A
                      common solution is the use of protecting cap layers on top
                      of the films. However they might change the physical
                      properties of the sample.Our solution is a handy mini
                      UHV-chamber which is used for sample transfer and quasi
                      in-situ measurements at the neutron reflectivity instrument
                      MARIA of the Jülich Centre for Neutron Science at MLZ in
                      Garching. The samples are prepared in the adjacent thin film
                      laboratory by molecular beam epitaxy and moved into the
                      compact chamber for transfer. It is equipped with sapphire
                      windows, an SAES getter pump and a wobble stick, which
                      serves also as a sample holder for samples of up to
                      1.4~cm$^2$. The pressure in the transfer chamber is kept
                      below $10^{-9}$ mbar. Neutron reflectivity measurements can
                      be performed at room temperature in magnetic fields of up to
                      600~mT. We present first polarized neutron reflectivity
                      measurements on Co thin films at room temperature in a
                      magnetic field of 300 mT in the Q-range up to 0.2
                      \AA$^{-1}$.},
      month         = {Mar},
      date          = {2016-03-06},
      organization  = {Frühjahrstagung der Deutschen
                       Physikalischen Gesellschaft, Regensburg
                       (Germany), 6 Mar 2016 - 11 Mar 2016},
      subtyp        = {After Call},
      cin          = {JCNS (München) ; Jülich Centre for Neutron Science JCNS
                      (München) ; JCNS-FRM-II / JCNS-2},
      cid          = {I:(DE-Juel1)JCNS-FRM-II-20110218 /
                      I:(DE-Juel1)JCNS-2-20110106},
      pnm          = {524 - Controlling Collective States (POF3-524) / 6212 -
                      Quantum Condensed Matter: Magnetism, Superconductivity
                      (POF3-621) / 6G15 - FRM II / MLZ (POF3-6G15) / 6G4 - Jülich
                      Centre for Neutron Research (JCNS) (POF3-623)},
      pid          = {G:(DE-HGF)POF3-524 / G:(DE-HGF)POF3-6212 /
                      G:(DE-HGF)POF3-6G15 / G:(DE-HGF)POF3-6G4},
      experiment   = {EXP:(DE-MLZ)MARIA-20140101 / EXP:(DE-MLZ)MBE-MLZ-20151210},
      typ          = {PUB:(DE-HGF)24},
      url          = {https://juser.fz-juelich.de/record/811981},
}