%0 Journal Article
%A Herzkamp, Matthias
%A Durini, Daniel
%A Degenhardt, Carsten
%A Erven, Andreas
%A Nöldgen, Holger
%A Feoktystov, Artem
%A Jokhovets, Lioubov
%A Streun, Matthias
%A Schwaitzer, Andreas
%A van Waasen, Stefan
%T Development and characterization of a 4 × 4 mm² pixel neutron scintillation detector using digital SiPMs
%J Journal of Instrumentation
%V 12
%N 12
%@ 1748-0221
%C London
%I Inst. of Physics
%M FZJ-2017-07997
%P C12001
%D 2017
%X This work describes the development of the first demonstrator device for neutrondetection based on a 6 Li-glass as a scintillator and silicon photomultipliers (SiPM) as photodetector.For the first characterization, the scintillator was pixelated with a one to one correspondence betweenscintillator and SiPM pixels, and optical cross-talk between pixels was minimized. Measurementsin a high luminosity neutron beam show the functionality of the device and allow for partialcharacterization. The position resolution is 4 ×4 mm² and the detection efficiency of neutrons is91(6) % relative to the active area. The device is linear up to at least 600 kcps.
%B 19th International Workshop on Radiation Imaging Detectors
%C 2 Jul 2017 - 6 Jul 2017, Krakow (Poland)
Y2 2 Jul 2017 - 6 Jul 2017
M2 Krakow, Poland
%F PUB:(DE-HGF)16
%9 Journal Article
%U https://juser.fz-juelich.de/record/840483