%0 Journal Article
%A Voigtländer, Bert
%A Coenen, Peter
%A Cherepanov, Vasily
%A Borgens, Peter
%A Duden, Thomas
%A Tautz, F. Stefan
%T Erratum: “Low vibration laboratory with a single-stage vibration isolation for microscopy applications” [Rev. Sci. Instrum. 88, 023703 (2017)]
%J Review of scientific instruments
%V 89
%N 1
%@ 1089-7623
%C [S.l.]
%I American Institute of Physics
%M FZJ-2018-00180
%P 019901-1 - 019901-9
%D 2018
%X Original article: Review of Scientific Instruments 88 (2), 023703 (2017)In our publication entitled “Low vibration laboratory with a single-stage vibration isolation for microscopy applications”1 we have erroneously labeled the measured spectrum Nspec of the vibrations as the spectral density NPSD ≡ Nvof the vibrations in our laboratory . Due to this, the measured vibration levels (spectral densities) shown in Figs. 3 and 5–7 have to be multiplied by a factor of 5.9. The corrected figures are shown in Figs. 1–5. Note that the magenta curve in Fig. 3 of Ref. 1 is not shifted because it was measured with a bandwidth of 1 Hz.
%F PUB:(DE-HGF)16
%9 Journal Article
%U <Go to ISI:>//WOS:000424022400084
%$ pmid:29390703
%R 10.1063/1.5008390
%U https://juser.fz-juelich.de/record/841883