TY - JOUR
AU - Voigtländer, Bert
AU - Coenen, Peter
AU - Cherepanov, Vasily
AU - Borgens, Peter
AU - Duden, Thomas
AU - Tautz, F. Stefan
TI - Erratum: “Low vibration laboratory with a single-stage vibration isolation for microscopy applications” [Rev. Sci. Instrum. 88, 023703 (2017)]
JO - Review of scientific instruments
VL - 89
IS - 1
SN - 1089-7623
CY - [S.l.]
PB - American Institute of Physics
M1 - FZJ-2018-00180
SP - 019901-1 - 019901-9
PY - 2018
AB - Original article: Review of Scientific Instruments 88 (2), 023703 (2017)In our publication entitled “Low vibration laboratory with a single-stage vibration isolation for microscopy applications”1 we have erroneously labeled the measured spectrum Nspec of the vibrations as the spectral density NPSD ≡ Nvof the vibrations in our laboratory . Due to this, the measured vibration levels (spectral densities) shown in Figs. 3 and 5–7 have to be multiplied by a factor of 5.9. The corrected figures are shown in Figs. 1–5. Note that the magenta curve in Fig. 3 of Ref. 1 is not shifted because it was measured with a bandwidth of 1 Hz.
LB - PUB:(DE-HGF)16
UR - <Go to ISI:>//WOS:000424022400084
C6 - pmid:29390703
DO - DOI:10.1063/1.5008390
UR - https://juser.fz-juelich.de/record/841883
ER -