@ARTICLE{Voigtlnder:841883,
author = {Voigtländer, Bert and Coenen, Peter and Cherepanov, Vasily
and Borgens, Peter and Duden, Thomas and Tautz, F. Stefan},
title = {{E}rratum: “{L}ow vibration laboratory with a
single-stage vibration isolation for microscopy
applications” [{R}ev. {S}ci. {I}nstrum. 88, 023703
(2017)]},
journal = {Review of scientific instruments},
volume = {89},
number = {1},
issn = {1089-7623},
address = {[S.l.]},
publisher = {American Institute of Physics},
reportid = {FZJ-2018-00180},
pages = {019901-1 - 019901-9},
year = {2018},
abstract = {Original article: Review of Scientific Instruments 88 (2),
023703 (2017)In our publication entitled “Low vibration
laboratory with a single-stage vibration isolation for
microscopy applications”1 we have erroneously labeled the
measured spectrum Nspec of the vibrations as the spectral
density NPSD ≡ Nvof the vibrations in our laboratory . Due
to this, the measured vibration levels (spectral densities)
shown in Figs. 3 and 5–7 have to be multiplied by a factor
of 5.9. The corrected figures are shown in Figs. 1–5. Note
that the magenta curve in Fig. 3 of Ref. 1 is not shifted
because it was measured with a bandwidth of 1 Hz.},
cin = {PGI-3 / JARA-FIT},
ddc = {530},
cid = {I:(DE-Juel1)PGI-3-20110106 / $I:(DE-82)080009_20140620$},
pnm = {141 - Controlling Electron Charge-Based Phenomena
(POF3-141)},
pid = {G:(DE-HGF)POF3-141},
typ = {PUB:(DE-HGF)16},
UT = {WOS:000424022400084},
pubmed = {pmid:29390703},
doi = {10.1063/1.5008390},
url = {https://juser.fz-juelich.de/record/841883},
}