%0 Journal Article
%A Rofors, E.
%A Perrey, H.
%A Al Jebali, R.
%A Annand, J. R. M.
%A Boyd, L.
%A Clemens, U.
%A Desert, S.
%A Engels, R.
%A Fissum, K. G.
%A Frielinghaus, H.
%A Gheorghe, C.
%A Hall-Wilton, R.
%A Jaksch, S.
%A Jalgén, A.
%A Kanaki, K.
%A Kemmerling, G.
%A Maulerova, V.
%A Mauritzson, N.
%A Montgomery, R.
%A Scherzinger, J.
%A Seitz, B.
%T Response of a Li-glass/multi-anode photomultiplier detector to α -particles from 241Am
%J Nuclear instruments & methods in physics research / A Accelerators, spectrometers, detectors and associated equipment Section A
%V 929
%@ 0168-9002
%C Amsterdam
%I North-Holland Publ. Co.
%M FZJ-2019-05964
%P 90 - 96
%D 2019
%X The response of a position-sensitive Li-glass scintillator detector to 𝛼-particles from a collimated 241Am source scanned across the face of the detector has been measured. Scintillation light was read out by an 8 × 8 pixel multi-anode photomultiplier and the signal amplitude for each pixel has been recorded for every position on a scan. The pixel signal is strongly dependent on position and in general several pixels will register a signal (a hit) above a given threshold. The effect of this threshold on hit multiplicity is studied, with a view to optimize the single-hit efficiency of the detector.
%F PUB:(DE-HGF)16
%9 Journal Article
%U <Go to ISI:>//WOS:000464347000011
%R 10.1016/j.nima.2019.03.014
%U https://juser.fz-juelich.de/record/866907