      title        = {{R}esponse of a {L}i-glass/multi-anode photomultiplier
                      detector to α -particles from 241{A}m},
      journal      = {Nuclear instruments $\&$ methods in physics research / A
                      Accelerators, spectrometers, detectors and associated
                      equipment Section A},
      volume       = {929},
      issn         = {0168-9002},
      address      = {Amsterdam},
      publisher    = {North-Holland Publ. Co.},
      reportid     = {FZJ-2019-05964},
      pages        = {90 - 96},
      year         = {2019},
      abstract     = {The response of a position-sensitive Li-glass scintillator
                      detector to 𝛼-particles from a collimated 241Am source
                      scanned across the face of the detector has been measured.
                      Scintillation light was read out by an 8 × 8 pixel
                      multi-anode photomultiplier and the signal amplitude for
                      each pixel has been recorded for every position on a scan.
                      The pixel signal is strongly dependent on position and in
                      general several pixels will register a signal (a hit) above
                      a given threshold. The effect of this threshold on hit
                      multiplicity is studied, with a view to optimize the
                      single-hit efficiency of the detector.},
      cin          = {JCNS-FRM-II / JCNS-1 / MLZ},
      ddc          = {530},
      cid          = {I:(DE-Juel1)JCNS-FRM-II-20110218 /
                      I:(DE-Juel1)JCNS-1-20110106 / I:(DE-588b)4597118-3},
      pnm          = {6G4 - Jülich Centre for Neutron Research (JCNS) (POF3-623)
                      / 6G15 - FRM II / MLZ (POF3-6G15)},
      pid          = {G:(DE-HGF)POF3-6G4 / G:(DE-HGF)POF3-6G15},
      experiment   = {EXP:(DE-MLZ)NOSPEC-20140101},
      typ          = {PUB:(DE-HGF)16},
      UT           = {WOS:000464347000011},
      doi          = {10.1016/j.nima.2019.03.014},
      url          = {https://juser.fz-juelich.de/record/866907},
}