% IMPORTANT: The following is UTF-8 encoded. This means that in the presence
% of non-ASCII characters, it will not work with BibTeX 0.99 or older.
% Instead, you should use an up-to-date BibTeX implementation like “bibtex8” or
% “biber”.
@ARTICLE{Bussmann:39769,
author = {Bussmann, N. and Engels, R. and Fuss, L. and Kemmerling, F.
G. and Reinartz, R. and Langen, K. J. and Schelten, J. and
Ziemons, K.},
title = {{P}erformance measurement of a new high-resolution detector
system for {I}-131 thyroid studies},
journal = {IEEE transactions on nuclear science},
volume = {48},
issn = {0018-9499},
address = {New York, NY},
publisher = {IEEE},
reportid = {PreJuSER-39769},
pages = {625 - 627},
year = {2001},
note = {Record converted from VDB: 12.11.2012},
abstract = {A two-dimensional detector system for high-resolution
thyroid I-131 scintigraphy was developed. It has a sensitive
area of 4 x 4 cm(2) and consists of a lead collimator and an
array of 10 x 10 BGO crystals combined with a
position-sensitive photomultiplier. The spatial resolution
and the sensitivity of the detector have been measured and
compared to two commercially available gamma cameras. The
full-width half-maximum for the new system is 3.8 mm.
Furthermore, first patient measurements have been carried
out.},
keywords = {J (WoSType)},
cin = {ISG-2 / ZEL / IME},
ddc = {620},
cid = {I:(DE-Juel1)VDB42 / I:(DE-Juel1)ZEL-20090406 /
I:(DE-Juel1)VDB54},
pnm = {Ionen- und Lithographietechnik / On-Line-Datenverarbeitung
/ Signaltransduktion},
pid = {G:(DE-Juel1)FUEK66 / G:(DE-Juel1)FUEK101 /
G:(DE-Juel1)FUEK89},
shelfmark = {Engineering, Electrical $\&$ Electronic / Nuclear Science
$\&$ Technology},
typ = {PUB:(DE-HGF)16},
UT = {WOS:000170576300002},
doi = {10.1109/23.940136},
url = {https://juser.fz-juelich.de/record/39769},
}