Home > Publications database > Performance measurement of a new high-resolution detector system for I-131 thyroid studies |
Journal Article | PreJuSER-39769 |
; ; ; ; ; ; ;
2001
IEEE
New York, NY
This record in other databases:
Please use a persistent id in citations: http://hdl.handle.net/2128/2090 doi:10.1109/23.940136
Abstract: A two-dimensional detector system for high-resolution thyroid I-131 scintigraphy was developed. It has a sensitive area of 4 x 4 cm(2) and consists of a lead collimator and an array of 10 x 10 BGO crystals combined with a position-sensitive photomultiplier. The spatial resolution and the sensitivity of the detector have been measured and compared to two commercially available gamma cameras. The full-width half-maximum for the new system is 3.8 mm. Furthermore, first patient measurements have been carried out.
Keyword(s): J ; detectors (auto) ; I-131 (auto) ; scintigraphy (auto) ; thyroid scanner (auto)
![]() |