%0 Journal Article
%A Voigtländer, Bert
%A Coenen, Peter
%A Cherepanov, Vasily
%A Borgens, Peter
%A Duden, Thomas
%A Tautz, F. S.
%T Low vibration laboratory with a single-stage vibration isolation for microscopy applications
%J Review of scientific instruments
%V 88
%N 2
%@ 0034-6748
%C [S.l.]
%I American Institute of Physics
%M FZJ-2017-01879
%P 023703 -1 - 023703-7
%D 2017
%X The construction and the vibrational performance of a low vibration laboratory for microscopy applications comprising a 100 ton floating foundation supported by passive pneumatic isolators (air springs), which rest themselves on a 200 ton solid base plate, are discussed. The optimization of the air spring system leads to a vibration level on the floating floor below that induced by an acceleration of 10 ng for most frequencies. Additional acoustic and electromagnetic isolation is accomplished by a room-in-room concept.
%F PUB:(DE-HGF)16
%9 Journal Article
%U <Go to ISI:>//WOS:000395902700027
%$ pmid:28249528
%R 10.1063/1.4975832
%U https://juser.fz-juelich.de/record/827775