Abstract: Different types of Ag/AgCl reference electrodes have been realised by means of thin- and thick-film technique. For inner electrolyte, KCl-containing membranes have been deposited and different coatings have been used to protect the reference electrode from a fast leaching out of KCl. The stability of the potential of the reference electrodes without KCl-containing membranes in 3 M KCl was about 7 hours for thin- film electrodes and up to 90 hours for thick-film electrodes. The reference electrodes with membrane were stable for more than two months. The thick-film reference electrode was integrated with a thin- film pH sensor onto one chip and the system pH sensor/reference electrode was investigated by means of capacitance/voltage (C/V) and constant capacitance (ConCap) mode.

Keyword(s): J ; reference electrode (auto) ; thin-film technique (auto) ; thick-film technique (auto) ; "one chip" sensor (auto) ; pH sensor (auto)


Note: Record converted from VDB: 12.11.2012

Contributing Institute(s):
  1. Institut für Halbleiterschichten und Bauelemente (ISG-1)
  2. Institut für Bio- und Chemosensoren (ISG-2)
Research Program(s):
  1. Materialien, Prozesse und Bauelemente für die Mikro- und Nanoelektronik (I01)

Appears in the scientific report 2003
Database coverage:
OpenAccess
Click to display QR Code for this record

The record appears in these collections:
Document types > Articles > Journal Article
Institute Collections > PGI > PGI-9
Workflow collections > Public records
IBN > IBN-2
Publications database
Open Access

 Record created 2012-11-13, last modified 2020-04-23