Abstract: Original article: Review of Scientific Instruments 88 (2), 023703 (2017)In our publication entitled “Low vibration laboratory with a single-stage vibration isolation for microscopy applications”1 we have erroneously labeled the measured spectrum Nspec of the vibrations as the spectral density NPSD ≡ Nvof the vibrations in our laboratory . Due to this, the measured vibration levels (spectral densities) shown in Figs. 3 and 5–7 have to be multiplied by a factor of 5.9. The corrected figures are shown in Figs. 1–5. Note that the magenta curve in Fig. 3 of Ref. 1 is not shifted because it was measured with a bandwidth of 1 Hz.
![]() |